Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Tek 2445: (1 year) CAL VER /5520+SC600 DATE: 08-Oct-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 521 NUMBER OF LINES: 3889 CONFIGURATION: Fluke 5520A (S6) ============================================================================= # # Source: # Tektronix 2445 Service Manual, April 1983 # # Compatibility: # 5520/CAL or MET/CAL or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N J F W 1.002 ASK+ D 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 20 min. - REQUIRED 1.004 DISP Temperature : 20 - 30 degC 1.004 DISP Humidity : <80% 1.005 HEAD [32] WARNING! 1.006 DISP DURING THE FOLLOWING PROCEDURE, HIGH VOLTAGES THAT COULD 1.006 DISP CAUSE SERIOUS INJURY OR EVEN DEATH ARE PRESENT! 1.006 DISP 1.006 DISP USE CAUTION TO PREVENT INJURY! 1.007 DISP This procedure requires the following... 1.007 DISP 1.007 DISP Function Generator; rep rate 1 kHz - 1 MHz, 15 VP-P 1.007 DISP Test Oscilloscope; 150 MHz Bandwidth 1.008 HEAD INITIAL CONTROL SETTINGS 1.009 DISP VERTICAL MODE : CH 1 ON, all others OFF 1.009 DISP CHOP/ALT : ALT (button out) 1.009 DISP 20 MHz BW LIMIT : Off (button out) 1.009 DISP CH 1 & CH 2 VOLTS/DIV : 1 V 1.009 DISP CH 1 & CH 2 VAR VOLTS/DIV : CAL (in detent) 1.009 DISP CH 3 & CH 4 VOLTS/DIV : 0.1 V (buttons out) 1.009 DISP CH 1 & CH 2 COUPLING : 1 MOhm GND 1.010 DISP A and B SEC/DIV : 10 ms (knobs locked) 1.010 DISP A and B SEC/DIV VAR : CAL (In detent) 1.010 DISP X10 MAG : Off (button out) 1.010 DISP Delta T and Delta V : Off (no readout) 1.010 DISP TRACKING : Off (button out) 1.010 DISP TRACE SEP : Fully CW 1.010 DISP TRIG HOLDOFF : Fully CCW 1.010 DISP TRIG LEVEL : Midrange 1.011 DISP TRIG SLOPE : + (plus) 1.011 DISP A/B TRIG SELECT : A 1.011 DISP TRIG MODE : AUTO LVL 1.011 DISP TRIG SOURCE : VERT 1.011 DISP TRIG COUPLING : DC 1.012 HEAD {CHANNEL 1 50 Ohm OVERLOAD PROTECTION} 1.013 DISP Connect the 5520A SCOPE BNC to the CH 1 OR 1.013 DISP X input connector via a 50 Ohm BNC cable. 1.014 DISP Change the CH 1 Input Coupling switch to 50 Ohm DC. 1.015 M5520 LIMIT 1T 1.016 5520 5V OM N 2W 1.017 EVAL -e MEM == 0 : No overload expected 2.001 DISP Change the CH 1 VOLTS/DIV control to 5 V/div. 2.002 M5520 LIMIT 15T 2.003 5520 9V OM N 2W 2.004 EVAL -e MEM : Overload expected 3.001 DISP Clear the OVERLOAD condition by pressing 3.001 DISP up on the CH 1 Input Coupling switch. 3.002 DISP Verify the CH 1, 1 MOhm DC indicator is lit 3.002 DISP and the readout display no longer indicates 3.002 DISP "50 Ohm OVERLOAD". 3.003 EVAL Does {50 Ohm overload protection reset properly}? 4.001 HEAD {CHANNEL 2 50 Ohm OVERLOAD PROTECTION} 4.002 DISP VERTICAL MODE : CH 2 4.003 DISP Connect the 5520A SCOPE BNC to the CH 2 input connector 4.003 DISP via a 50 Ohm BNC cable. 4.004 DISP Change the CH 2 Input Coupling switch to 50 Ohm DC. 4.005 M5520 LIMIT 1T 4.006 5520 5V OM N 2W 4.007 EVAL -e MEM == 0 : No overload expected 5.001 DISP Change the CH 2 VOLTS/DIV control to 5 V/div. 5.002 M5520 LIMIT 15T 5.003 5520 9V OM N 2W 5.004 EVAL -e MEM : Overload expected 6.001 DISP Clear the OVERLOAD condition by pressing 6.001 DISP up on the CH 2 Input Coupling switch. 6.002 DISP Verify the CH 2, 1 MOhm DC indicator is lit 6.002 DISP and the readout display no longer indicates 6.002 DISP "50 Ohm OVERLOAD". 6.003 EVAL Does {50 Ohm overload protection reset properly}? 7.001 M5520 * 7.002 HEAD {CHANNEL 1 LOW-FREQUENCY AC COUPLING} 7.003 DISP VERT MODE : CH 1 7.003 DISP A and B SEC/DIV : 10 ms (knobs locked) 7.003 DISP CH 1 & CH 2 VOLTS/DIV : 10 mV 7.003 DISP CH 1 & CH 2 COUPLING : 1 MOhm GND 7.004 DISP NOTE: Prior to performing the following steps, the 7.004 DISP 10X probe must be properly compensated. 7.005 DISP Connect the Tektronix 2445 CALIBRATOR output signal 7.005 DISP to the CH 1 input connector using a 10X probe. 7.006 DISP Position the ground-reference trace 2 divisions 7.006 DISP below the center horizontal graticule line. 7.007 DISP CH 1 COUPLING : 1 MOhm DC 7.007 DISP 7.007 DISP Check that displayed signal is vertically centered 7.007 DISP and has an amplitude of 3.88 to 4.12 divisions. 7.008 EVAL Is {display vertically centered and 3.88 to 4.12 divisions}? 8.001 DISP CH 1 COUPLING : upper 1 MOhm GND 8.001 DISP 8.001 DISP Using the CH 1 POSITION control, align the 8.001 DISP trace with the center horizontal graticule line. 8.002 DISP Set the CH 1 Input Coupling switch to 1 MOhm AC. 8.002 DISP 8.002 DISP Check that displayed signal is a tilted square 8.002 DISP wave, 4.5 to 4.9 divisions in amplitude, 8.002 DISP vertically centered on the graticule. 8.003 EVAL Is {signal tilted, centered, and 4.5 to 4.9 div in amplitude}? 9.001 HEAD {CHANNEL 2 LOW-FREQUENCY AC COUPLING} 9.002 DISP Connect the Tektronix 2445 CALIBRATOR output signal 9.002 DISP to the CH 2 input connector using a 10X probe. 9.002 DISP 9.003 DISP Set the VERTICAL MODE switches to deselect CH 1 9.003 DISP and display CH 2. 9.003 DISP 9.004 DISP Position the ground-reference trace 2 divisions 9.004 DISP below the center horizontal graticule line. 9.005 DISP VERT MODE : CH 2 9.005 DISP CH 2 COUPLING : 1 MOhm DC 9.005 DISP 9.005 DISP Check that displayed signal is vertically centered 9.005 DISP and has an amplitude of 3.88 to 4.12 divisions. 9.006 EVAL Is {display vertically centered and 3.88 to 4.12 divisions}? 10.001 DISP CH 2 COUPLING : upper 1 MOhm GND 10.001 DISP 10.001 DISP Using the CH 2 POSITION control, align the 10.001 DISP trace with the center horizontal graticule line. 10.002 DISP Set the CH 2 Input Coupling switch to 1 MOhm AC. 10.002 DISP 10.002 DISP Check that displayed signal is a tilted square 10.002 DISP wave, 4.5 to 4.9 divisions in amplitude, 10.002 DISP vertically centered on the graticule. 10.003 EVAL Is {signal tilted, centered, and 4.5 to 4.9 div in amplitude}? 11.001 DISP Disconnect the 10X probe from the UUT 11.001 DISP 11.002 HEAD {CH 1 VOLTS/DIV, DELTA V, & TRIG LVL READOUT} 11.003 DISP CH 1 & CH 2 COUPLING : 1 MOhm AC 11.003 DISP VERT MODE : CH 1 11.003 DISP BW LIMIT : On (button in) 11.003 DISP Delta V : On 11.003 DISP CH 1 & CH 2 VOLTS/DIV : 2 mV 11.003 DISP A and B SEC/DIV : 1 ms (knobs locked) 11.003 DISP TRIGGER MODE : AUTO 11.004 DISP Press up and momentarily hold the CH 1 and CH 2 11.004 DISP Input Coupling switches in their 1 MOhm AC 11.004 DISP positions until a moving dot display 11.004 DISP replaces the normal signal and readout 11.004 DISP displays (for approximately 15 seconds). 11.004 DISP This performs a DC Balance of CH 1 and CH 2. 11.005 DISP When the signal and readout displays 11.005 DISP automatically return to normal, set the CH 1 11.005 DISP and CH 2 Input Coupling to 1 MOhm DC. 11.005 DISP 11.006 DISP MAKE THE FOLLOWING CONNECTION... 11.006 DISP 11.006 DISP 5520 SCOPE OUTPUT --------- CH 1 INPUT 11.006 DISP 11.007 RSLT =VOLTS/DIV ACCURACY: 2 mV/DIV 11.008 ASK+ N 11.009 5520 2 10mVpp 2% 1kH ZQ S6 12.001 HEAD {DELTA VOLTS READOUT} 12.002 DISP Press and release the Delta V pushbutton to 12.002 DISP obtain the Delta V readout display. 12.003 5520 10mVpp 1kH ZQ S6 S 12.004 DISP Use the VERTICAL POSITION control to set the 12.004 DISP bottom of the signal 2 divisions below graticule 12.004 DISP center. 12.004 DISP 12.005 DISP Throughout the DELTA VOLTS READOUT tests... 12.005 DISP 12.005 DISP Rotate the Delta REF OR DLY POS control to 12.005 DISP align the reference cursor with the bottom of 12.005 DISP the waveform. 12.005 DISP 12.005 DISP Rotate the Delta control to align the Delta cursor 12.005 DISP with the top of the signal display. 12.006 EVAL Is {delta volts readout 9.81 to 10.20 mV}? 13.001 HEAD {A TRIGGER LEVEL: +Peak} 13.002 5520 10mVpp 1kH ZQ S6 S 13.003 DISP Throughout the A TRIGGER LEVEL +PEAK tests... 13.003 DISP 13.003 DISP Set the TRIGGER LEVEL control at the 13.003 DISP most positive voltage that produces a 13.003 DISP barely triggered, jittering display. 13.003 DISP 13.004 EVAL Is {positive trigger level readout 8.5 to 11.5 mV}? 14.001 HEAD A TRIGGER LEVEL: {-Peak} 14.002 DISP TRIGGER SLOPE : - 14.002 DISP 14.003 5520 10mVpp 1kH ZQ S6 S 14.004 DISP Throughout the A TRIGGER LEVEL -PEAK tests... 14.004 DISP 14.004 DISP Set the TRIGGER LEVEL control at the 14.004 DISP most negative voltage that produces a 14.004 DISP barely triggered, jittering display. 14.004 DISP 14.005 EVAL Is {negative trigger level readout +1.2 to -1.2 mV}? 15.001 ASK- J F W 15.002 ASK+ R D N 15.003 HEAD {B TRIGGER LEVEL: +Peak} 15.004 DISP TRIGGER SLOPE : + 15.004 DISP 15.005 5520 10mVpp 1kH ZQ S6 S 15.006 DISP Set the TRIGGER LEVEL for a stable display. 15.006 DISP 15.006 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 15.006 DISP to TRIG AFT DLY. 15.007 DISP Adjust Delta REF OR DLY POS control for a 15.007 DISP delay readout of 0.000 ms. 15.008 DISP Set the TRIGGER LEVEL control to the 15.008 DISP most positive voltage that produces an 15.008 DISP intensified point on the waveform display. 15.008 DISP 15.009 EVAL Is {positive trigger level readout 8.5 to 11.5 mV}? 16.001 HEAD B TRIGGER LEVEL: {-Peak} 16.002 DISP TRIGGER SLOPE : - 16.002 DISP 16.003 5520 10mVpp 1kH ZQ S6 S 16.004 DISP Set the TRIGGER LEVEL control to the 16.004 DISP most negative voltage that produces an 16.004 DISP intensified point on the waveform display. 16.004 DISP 16.005 EVAL Is {negative trigger level readout +1.2 to -1.2 mV}? 17.001 DISP Push the B SEC/DIV knob in. 17.002 HEAD {VOLTS/DIV ACCURACY: 5 mV/DIV} 17.003 ASK+ R 17.004 DISP Press and release the Delta V pushbutton to 17.004 DISP reobtain the Delta V readout display. 17.005 DISP TRIGGER SLOPE : + 17.005 DISP 17.006 5520 5 20mVpp 2% 1kH ZQ S6 18.001 HEAD {DELTA VOLTS READOUT} 18.002 5520 20mVpp 1kH ZQ S6 S 18.003 EVAL Is {delta volts readout 19.6 to 20.4 mV}? 19.001 HEAD {A TRIGGER LEVEL: +Peak} 19.002 5520 20mVpp 1kH ZQ S6 S 19.003 EVAL Is {positive trigger level readout 17.3 to 22.7 mV}? 20.001 HEAD A TRIGGER LEVEL: {-Peak} 20.002 DISP TRIGGER SLOPE : - 20.002 DISP 20.003 5520 20mVpp 1kH ZQ S6 S 20.004 EVAL Is {negative trigger level readout +2.1 to -2.1 mV}? 21.001 HEAD {B TRIGGER LEVEL: +Peak} 21.002 DISP TRIGGER SLOPE : + 21.002 DISP 21.003 5520 20mVpp 1kH ZQ S6 S 21.004 DISP Set the TRIGGER LEVEL for a stable display. 21.004 DISP 21.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 21.004 DISP to TRIG AFT DLY. 21.005 EVAL Is {positive trigger level readout 17.3 to 22.7 mV}? 22.001 HEAD B TRIGGER LEVEL: {-Peak} 22.002 DISP TRIGGER SLOPE : - 22.002 DISP 22.003 5520 20mVpp 1kH ZQ S6 S 22.004 EVAL Is {negative trigger level readout +2.1 to -2.1 mV}? 23.001 DISP Push the B SEC/DIV knob in. 23.002 HEAD {VOLTS/DIV ACCURACY: 10 mV/DIV} 23.003 DISP Press and release the Delta V pushbutton to 23.003 DISP reobtain the Delta V readout display. 23.004 DISP TRIGGER SLOPE : + 23.004 DISP 23.005 5520 10 50mVpp 2% 1kH ZQ S6 24.001 HEAD {DELTA VOLTS READOUT} 24.002 5520 50mVpp 1kH ZQ S6 S 24.003 EVAL Is {delta volts readout 49.0 to 50.9 mV}? 25.001 HEAD {A TRIGGER LEVEL: +Peak} 25.002 5520 50mVpp 1kH ZQ S6 S 25.003 EVAL Is {positive trigger level readout 44.5 to 55.5 mV}? 26.001 HEAD A TRIGGER LEVEL: {-Peak} 26.002 DISP TRIGGER SLOPE : - 26.002 DISP 26.003 5520 50mVpp 1kH ZQ S6 S 26.004 EVAL Is {negative trigger level readout +4 to -4 mV}? 27.001 HEAD {B TRIGGER LEVEL: +Peak} 27.002 DISP TRIGGER SLOPE : + 27.002 DISP 27.003 5520 50mVpp 1kH ZQ S6 S 27.004 DISP Set the TRIGGER LEVEL for a stable display. 27.004 DISP 27.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 27.004 DISP to TRIG AFT DLY. 27.005 EVAL Is {positive trigger level readout 44.5 to 55.5 mV}? 28.001 HEAD B TRIGGER LEVEL: {-Peak} 28.002 DISP TRIGGER SLOPE : - 28.002 DISP 28.003 5520 50mVpp 1kH ZQ S6 S 28.004 EVAL Is {negative trigger level readout +4 to -4 mV}? 29.001 DISP Push the B SEC/DIV knob in. 29.002 HEAD {VOLTS/DIV ACCURACY: 20 mV/DIV} 29.003 DISP Press and release the Delta V pushbutton to 29.003 DISP reobtain the Delta V readout display. 29.004 DISP TRIGGER SLOPE : + 29.004 DISP 29.005 5520 20 100mVpp 2% 1kH ZQ S6 30.001 HEAD {DELTA VOLTS READOUT} 30.002 5520 100mVpp 1kH ZQ S6 S 30.003 EVAL Is {delta volts readout 98.1 to 102.0 mV}? 31.001 HEAD {A TRIGGER LEVEL: +Peak} 31.002 5520 100mVpp 1kH ZQ S6 S 31.003 EVAL Is {positive trigger level readout 89 to 111 mV}? 32.001 HEAD A TRIGGER LEVEL: {-Peak} 32.002 DISP TRIGGER SLOPE : - 32.002 DISP 32.003 5520 100mVpp 1kH ZQ S6 S 32.004 EVAL Is {negative trigger level readout +7.5 to -7.5 mV}? 33.001 HEAD {B TRIGGER LEVEL: +Peak} 33.002 DISP TRIGGER SLOPE : + 33.002 DISP 33.003 5520 100mVpp 1kH ZQ S6 S 33.004 DISP Set the TRIGGER LEVEL for a stable display. 33.004 DISP 33.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 33.004 DISP to TRIG AFT DLY. 33.005 EVAL Is {positive trigger level readout 89 to 111 mV}? 34.001 HEAD B TRIGGER LEVEL: {-Peak} 34.002 DISP TRIGGER SLOPE : - 34.002 DISP 34.003 5520 100mVpp 1kH ZQ S6 S 34.004 EVAL Is {negative trigger level readout +7.5 to -7.5 mV}? 35.001 DISP Push the B SEC/DIV knob in. 35.002 HEAD {VOLTS/DIV ACCURACY: 50 mV/DIV} 35.003 DISP Press and release the Delta V pushbutton to 35.003 DISP reobtain the Delta V readout display. 35.004 DISP TRIGGER SLOPE : + 35.004 DISP 35.005 5520 50 200mVpp 2% 1kH ZQ S6 36.001 HEAD {DELTA VOLTS READOUT} 36.002 5520 200mVpp 1kH ZQ S6 S 36.003 EVAL Is {delta volts readout 196 to 204 mV}? 37.001 HEAD {A TRIGGER LEVEL: +Peak} 37.002 5520 200mVpp 1kH ZQ S6 S 37.003 EVAL Is {positive trigger level readout 177 to 223 mV}? 38.001 HEAD A TRIGGER LEVEL: {-Peak} 38.002 DISP TRIGGER SLOPE : - 38.002 DISP 38.003 5520 200mVpp 1kH ZQ S6 S 38.004 EVAL Is {negative trigger level readout +17 to -17 mV}? 39.001 HEAD {NOISE REJ: +Peak} 39.002 DISP TRIGGER SLOPE : + 39.002 DISP 39.003 DISP TRIGGER COUPLING : NOISE REJ 39.004 5520 200mVpp 1kH ZQ S6 S 39.005 EVAL Is {positive trigger level readout 147 to 253 mV}? 40.001 HEAD {NOISE REJ: -Peak} 40.002 DISP TRIGGER SLOPE : - 40.002 DISP 40.003 5520 200mVpp 1kH ZQ S6 S 40.004 EVAL Is {negative trigger level readout +47 to -47 mV}? 41.001 HEAD {B TRIGGER LEVEL: +Peak} 41.002 DISP TRIGGER SLOPE : + 41.002 DISP 41.003 DISP TRIGGER COUPLING : DC 41.004 5520 200mVpp 1kH ZQ S6 S 41.005 DISP Set the TRIGGER LEVEL for a stable display. 41.005 DISP 41.005 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 41.005 DISP to TRIG AFT DLY. 41.006 EVAL Is {positive trigger level readout 177 to 223 mV}? 42.001 HEAD B TRIGGER LEVEL: {-Peak} 42.002 DISP TRIGGER SLOPE : - 42.002 DISP 42.003 5520 200mVpp 1kH ZQ S6 S 42.004 EVAL Is {negative trigger level readout +17 to -17 mV}? 43.001 DISP Push the B SEC/DIV knob in. 43.002 HEAD {VOLTS/DIV ACCURACY: .1 V/DIV} 43.003 DISP Press and release the Delta V pushbutton to 43.003 DISP reobtain the Delta V readout display. 43.004 DISP TRIGGER SLOPE : + 43.004 DISP 43.005 5520 0.1 0.5Vpp 2% 1kH ZQ S6 44.001 HEAD {DELTA VOLTS READOUT} 44.002 5520 0.5Vpp 1kH ZQ S6 S 44.003 EVAL Is {delta volts readout 490 to 509 mV}? 45.001 HEAD {A TRIGGER LEVEL: +Peak} 45.002 5520 0.5Vpp 1kH ZQ S6 S 45.003 EVAL Is {positive trigger level readout 0.449 to 0.551 V}? 46.001 HEAD A TRIGGER LEVEL: {-Peak} 46.002 DISP TRIGGER SLOPE : - 46.002 DISP 46.003 5520 0.5Vpp 1kH ZQ S6 S 46.004 EVAL Is {negative trigger level readout +0.036 to -0.036 V}? 47.001 HEAD {B TRIGGER LEVEL: +Peak} 47.002 DISP TRIGGER SLOPE : + 47.002 DISP 47.003 5520 0.5Vpp 1kH ZQ S6 S 47.004 DISP Set the TRIGGER LEVEL for a stable display. 47.004 DISP 47.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 47.004 DISP to TRIG AFT DLY. 47.005 EVAL Is {positive trigger level readout 0.449 to 0.551 V}? 48.001 HEAD B TRIGGER LEVEL: {-Peak} 48.002 DISP TRIGGER SLOPE : - 48.002 DISP 48.003 5520 0.5Vpp 1kH ZQ S6 S 48.004 EVAL Is {negative trigger level readout +0.036 to -0.036 V}? 49.001 DISP Push the B SEC/DIV knob in. 49.002 HEAD {VOLTS/DIV ACCURACY: .2 V/DIV} 49.003 DISP Press and release the Delta V pushbutton to 49.003 DISP reobtain the Delta V readout display. 49.004 DISP TRIGGER SLOPE : + 49.004 DISP 49.005 5520 0.2 1Vpp 2% 1kH ZQ S6 50.001 HEAD {DELTA VOLTS READOUT} 50.002 5520 1Vpp 1kH ZQ S6 S 50.003 EVAL Is {delta volts readout 0.981 to 1.020 V}? 51.001 HEAD {A TRIGGER LEVEL: +Peak} 51.002 5520 1Vpp 1kH ZQ S6 S 51.003 EVAL Is {positive trigger level readout 0.90 to 1.10 V}? 52.001 HEAD A TRIGGER LEVEL: {-Peak} 52.002 DISP TRIGGER SLOPE : - 52.002 DISP 52.003 5520 1Vpp 1kH ZQ S6 S 52.004 EVAL Is {negative trigger level readout +0.07 to -0.07 V}? 53.001 HEAD {B TRIGGER LEVEL: +Peak} 53.002 DISP TRIGGER SLOPE : + 53.002 DISP 53.003 5520 1Vpp 1kH ZQ S6 S 53.004 DISP Set the TRIGGER LEVEL for a stable display. 53.004 DISP 53.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 53.004 DISP to TRIG AFT DLY. 53.005 EVAL Is {positive trigger level readout 0.90 to 1.10 V}? 54.001 HEAD B TRIGGER LEVEL: {-Peak} 54.002 DISP TRIGGER SLOPE : - 54.002 DISP 54.003 5520 1Vpp 1kH ZQ S6 S 54.004 EVAL Is {negative trigger level readout +0.07 to -0.07 V}? 55.001 DISP Push the B SEC/DIV knob in. 55.002 HEAD {VOLTS/DIV ACCURACY: .5 V/DIV} 55.003 DISP Press and release the Delta V pushbutton to 55.003 DISP reobtain the Delta V readout display. 55.004 DISP TRIGGER SLOPE : + 55.004 DISP 55.005 5520 0.5 2Vpp 2% 1kH ZQ S6 56.001 HEAD {DELTA VOLTS READOUT} 56.002 5520 2Vpp 1kH ZQ S6 S 56.003 EVAL Is {delta volts readout 1.96 to 2.04 V}? 57.001 HEAD {A TRIGGER LEVEL: +Peak} 57.002 5520 2Vpp 1kH ZQ S6 S 57.003 EVAL Is {positive trigger level readout 1.78 to 2.22 V}? 58.001 HEAD A TRIGGER LEVEL: {-Peak} 58.002 DISP TRIGGER SLOPE : - 58.002 DISP 58.003 5520 2Vpp 1kH ZQ S6 S 58.004 EVAL Is {negative trigger level readout +0.14 to -0.14 V}? 59.001 HEAD {B TRIGGER LEVEL: +Peak} 59.002 DISP TRIGGER SLOPE : + 59.002 DISP 59.003 5520 2Vpp 1kH ZQ S6 S 59.004 DISP Set the TRIGGER LEVEL for a stable display. 59.004 DISP 59.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 59.004 DISP to TRIG AFT DLY. 59.005 EVAL Is {positive trigger level readout 1.78 to 2.22 V}? 60.001 HEAD B TRIGGER LEVEL: {-Peak} 60.002 DISP TRIGGER SLOPE : - 60.002 DISP 60.003 5520 2Vpp 1kH ZQ S6 S 60.004 EVAL Is {negative trigger level readout +0.14 to -0.14 V}? 61.001 DISP Push the B SEC/DIV knob in. 61.002 HEAD {VOLTS/DIV ACCURACY: 1 V/DIV} 61.003 DISP Press and release the Delta V pushbutton to 61.003 DISP reobtain the Delta V readout display. 61.004 DISP TRIGGER SLOPE : + 61.004 DISP 61.005 5520 1 5Vpp 2% 1kH ZQ S6 62.001 HEAD {DELTA VOLTS READOUT} 62.002 5520 5Vpp 1kH ZQ S6 S 62.003 EVAL Is {delta volts readout 4.90 to 5.09 V}? 63.001 HEAD {A TRIGGER LEVEL: +Peak} 63.002 5520 5Vpp 1kH ZQ S6 S 63.003 EVAL Is {positive trigger level readout 4.50 to 5.50 V}? 64.001 HEAD A TRIGGER LEVEL: {-Peak} 64.002 DISP TRIGGER SLOPE : - 64.002 DISP 64.003 5520 5Vpp 1kH ZQ S6 S 64.004 EVAL Is {negative trigger level readout +0.35 to -0.35 V}? 65.001 HEAD {B TRIGGER LEVEL: +Peak} 65.002 DISP TRIGGER SLOPE : + 65.002 DISP 65.003 5520 5Vpp 1kH ZQ S6 S 65.004 DISP Set the TRIGGER LEVEL for a stable display. 65.004 DISP 65.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 65.004 DISP to TRIG AFT DLY. 65.005 EVAL Is {positive trigger level readout 4.50 to 5.50 V}? 66.001 HEAD B TRIGGER LEVEL: {-Peak} 66.002 DISP TRIGGER SLOPE : - 66.002 DISP 66.003 5520 5Vpp 1kH ZQ S6 S 66.004 EVAL Is {negative trigger level readout +0.35 to -0.35 V}? 67.001 DISP Push the B SEC/DIV knob in. 67.002 HEAD {VOLTS/DIV ACCURACY: 2 V/DIV} 67.003 DISP Press and release the Delta V pushbutton to 67.003 DISP reobtain the Delta V readout display. 67.004 DISP TRIGGER SLOPE : + 67.004 DISP 67.005 5520 2 10Vpp 2% 1kH ZQ S6 68.001 HEAD {DELTA VOLTS READOUT} 68.002 5520 10Vpp 1kH ZQ S6 S 68.003 EVAL Is {delta volts readout 9.81 to 10.2 V}? 69.001 HEAD {A TRIGGER LEVEL: +Peak} 69.002 5520 10Vpp 1kH ZQ S6 S 69.003 EVAL Is {positive trigger level readout 9.0 to 11.0 V}? 70.001 HEAD A TRIGGER LEVEL: {-Peak} 70.002 DISP TRIGGER SLOPE : - 70.002 DISP 70.003 5520 10Vpp 1kH ZQ S6 S 70.004 EVAL Is {negative trigger level readout +0.7 to -0.7 V}? 71.001 HEAD {B TRIGGER LEVEL: +Peak} 71.002 DISP TRIGGER SLOPE : + 71.002 DISP 71.003 5520 10Vpp 1kH ZQ S6 S 71.004 DISP Set the TRIGGER LEVEL for a stable display. 71.004 DISP 71.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 71.004 DISP to TRIG AFT DLY. 71.005 EVAL Is {positive trigger level readout 9.0 to 11.0 V}? 72.001 HEAD B TRIGGER LEVEL: {-Peak} 72.002 DISP TRIGGER SLOPE : - 72.002 DISP 72.003 5520 10Vpp 1kH ZQ S6 S 72.004 EVAL Is {negative trigger level readout +0.7 to -0.7 V}? 73.001 DISP Push the B SEC/DIV knob in. 73.002 HEAD {VOLTS/DIV ACCURACY: 5 V/DIV} 73.003 DISP Press and release the Delta V pushbutton to 73.003 DISP reobtain the Delta V readout display. 73.004 DISP TRIGGER SLOPE : + 73.004 DISP 73.005 5520 5 20Vpp 2% 1kH ZQ S6 74.001 HEAD {DELTA VOLTS READOUT} 74.002 5520 20Vpp 1kH ZQ S6 S 74.003 EVAL Is {delta volts readout 19.6 to 20.4 V}? 75.001 HEAD {A TRIGGER LEVEL: +Peak} 75.002 5520 20Vpp 1kH ZQ S6 S 75.003 EVAL Is {positive trigger level readout 17.8 to 22.2 V}? 76.001 HEAD A TRIGGER LEVEL: {-Peak} 76.002 DISP TRIGGER SLOPE : - 76.002 DISP 76.003 5520 20Vpp 1kH ZQ S6 S 76.004 EVAL Is {negative trigger level readout +1.4 to -1.4 V}? 77.001 HEAD {B TRIGGER LEVEL: +Peak} 77.002 DISP TRIGGER SLOPE : + 77.002 DISP 77.003 5520 20Vpp 1kH ZQ S6 S 77.004 DISP Set the TRIGGER LEVEL for a stable display. 77.004 DISP 77.004 DISP Pull the SEC/DIV knob out and set the B TRIGGER MODE 77.004 DISP to TRIG AFT DLY. 77.005 EVAL Is {positive trigger level readout 17.8 to 22.2 V}? 78.001 HEAD B TRIGGER LEVEL: {-Peak} 78.002 DISP TRIGGER SLOPE : - 78.002 DISP 78.003 5520 20Vpp 1kH ZQ S6 S 78.004 EVAL Is {negative trigger level readout +1.4 to -1.4 V}? 79.001 DISP Push the B SEC/DIV knob in. 79.002 ASK- N J F W 79.003 ASK+ R D 79.004 HEAD {VARIABLE VOLTS/DIV} 79.005 5520 25Vpp 1kH ZQ S6 S 79.006 MESS Check that display amplitude reduces to 2 divisions 79.006 MESS or less when the CH 1 VARIABLE VOLTS/DIV control 79.006 MESS is rotated fully CCW. 79.006 MESS 79.007 EVAL Does {VAR VOLTS/DIV reduce amplitude to